11
12
STATIONARY CONFIGURATION
2
STATIONARY CONFIGURATION
2
POSITIONING AND CARRYING THE COMPRESSOR
Position the compressor upright on a flat, level surface. Make sure that the cooling vents, cooling filter cover, and
air inlet on the back of the compressor are not blocked, and there is sufficient clearance from surrounding objects.
Protect the compressor from falling.
The compressor should not be used adjacent to or stacked with other equipment and that if adjacent or stacked
use is necessary, the compressor should be observed to verify normal operation in the configuration in which it
will be used.
When carrying the compressor, make sure to use the handle, keep the compressor in an upright position, and
protect it from falling or dropping.
NOTES:
• The compressor outputs an audible sound and must be at least 3 feet (1 m) away from the user during use.
• When operating in a 40°C (104°F) environment, the gas output by the ventilation system may reach
temperatures up to 48°C (118.4°F). To reduce patient discomfort, operate the ventilation system in a cooler
environment.
• The performance of the compressor may degrade in high temperature, high humidity, or high altitude
environments. If degradation is seen, switch to an alternate means of ventilation. Verify the performance of
the compressor for adequate therapy delivery in the environment(s) in which it will be used and adjust the
volume to compensate for altitude when necessary.
• The ventilator settings might not be achieved
when sourced by the Life2000
®
Compressor due
to increases in altitude near or above 2500 feet.
Consult the table to ensure that the compressor
can meet the ventilator settings. The tidal volume
delivered to the patient includes the ventilator
set entrainment volume from patient
int supplemental oxygen volume (if used).
For additional information see page 150.
WARNING:
• Do not use the ventilation system in the presence of flammable anesthetics.
• Do not cover or block the compressor’s internal alarm buzzer with any object. Covering the buzzer may
make it difficult for a patient or caregiver to hear alarms, which may result in inadequate respiratory
therapy.
• Do not cover the ventilator, touch screen, speaker, or backup alarm buzzer with tape or any other object.
Covering the ventilator or any of its parts might cause difficulty in hearing alarms and might affect ventilator
performance.
• When the ventilator is in use, keep it in a well-ventilated area to prevent it from overheating. The ventilator
may overheat and be permanently damaged if it is used in an area that is not well ventilated.
• Do not connect the ventilation system components or accessories to any other equipment that is not
described in this
Instructions for Use
.
CAUTION:
• Keep in a clean environment to protect the ventilation system from ingress of dust, lint, and pests.
• Do not leave the ventilation system exposed to the sun or other sources of radiant heat, it may overheat.
• Do not allow children or pets to access the ventilation system; it may become damaged.
Simulated Elevation
(in feet)
Observed Maximum
Compressor Output
(in LPM)
0
17
2500
14
4000
12
8000
8
INTRODUCTION TO STATIONARY CONFIGURATION
In Stationary Configuration, the ventilator is docked into the compressor. The compressor provides the ventilator
with a continuous pressure source and functions as a charging station.
STATIONARY CONFIGURATION SETUP CHECKLIST
Refer to the chapter contents for full instructions, warnings, and cautions.
☐
Test the ventilator before using it on a new patient if in multi-patient environment (see section below).
☐
Position the compressor
☐
Connect the compressor to an AC power source using the compressor’s external power supply and AC
power cord
☐
Dock the ventilator into the compressor
TIP:
Check to make sure the ventilator is locked in the compressor.
☐
Connect an interface to the compressor
NOTE:
For information about the interface on the patient side, see "Chapter 5: Connecting an Interface" on
☐
Power on the compressor
☐
Power on the ventilator
☐
Check the compressor’s battery charge status
☐
Check the ventilator’s battery charge
NOTE:
Ventilation will not begin until an Activity button is pressed on the ventilator. For more information see
"Choosing an Activity Button (Patient Selectable) to Begin Ventilation" on page 67.
TESTING THE VENTILATION SYSTEM
In a multi-patient setting, the ventilation system must be tested before it is assigned to a new patient. For
instructions on testing the ventilation system, see "Testing Ventilator Alarms" on page 115.