EXTENDED RANGE CONFIGURATION
3
EXTENDED RANGE CONFIGURATION
3
CONNECTING AN INTERFACE TO THE VENTILATOR IN EXTENDED RANGE
CONFIGURATION
Plug the Breathe Pillows Entrainment Interface or Universal Circuit
®
Connector into the interface port on the bottom of the ventilator until it
clicks. For more information about wearing interfaces, see "Chapter 5:
Connecting an Interface" on page 44.
NOTE:
Ensure that the Breathe Pillows Entrainment Interface or Universal
Circuit
®
Connector is connected to the ventilator in Extended Range
Configuration and Stand-Alone Configuration.
THE COMBO
2
® HOSE
The CombO
2
® hose may be used to connect the Life2000® Ventilation System in Extended Range Configuration
and provide supplemental low-flow oxygen from an oxygen concentrator or other low-flow oxygen source.
WARNING:
The interface, source gas supply hose, and power cords should be positioned to avoid restricting movement,
causing a tripping hazard, or posing a strangulation risk.
CAUTION
Use only an approved source gas supply hose with the ventilation system. If an unauthorized source gas supply
hose is used with the ventilation system, the system may be damaged.
NOTES:
• Visually inspect the CombO
2
® hose before using it.
• During use in Extended Range Configuration, the source gas supply hose should remain connected to the
ventilator at all times, except when required to be disconnected for maintenance, testing, or replacement. If
it is disconnected while the ventilator is on and delivering therapy, a Low Gas Pressure alarm will occur.
Oxygen tubing connector
Oxygen tubing
Oxygen tubing Connector
Outlet fitting connector
Source Gas Supply hose
Quick connect end