ALARMS, ALERTS, AND TROUBLESHOOTING
7
ALARMS, ALERTS, AND TROUBLESHOOTING
7
MEDIUM-PRIORITY ALARMS
NOTIFICATION
CAUSE
CHECKS AND POSSIBLE RESOLUTION IN EACH CONFIGURATION
STAND-ALONE CONFIGURATION
EXTENDED RANGE CONFIGURATION
STATIONARY CONFIGURATION
Battery Low
10:11a, Fri
Jun 12, 2015
12
Breath/min
Air LPM
PIP cmH
2
O
3.9
1.1
Battery Low
Ventilator battery capacity drops
below 25%.
Connect the ventilator to the ventilator battery
charger and an AC power source to recharge the
battery.
OR
Dock the ventilator back into the compressor to
recharge the battery. Ensure that the locked icon
on the compressor illuminates to indicate the
ventilator is charging. Ensure the compressor is
powered on.
If the battery does not recharge, contact your
service representative.
Connect the ventilator to the ventilator battery
charger and an AC power source to recharge the
battery.
OR
Dock the ventilator back into the compressor to
recharge the battery. Ensure that the locked icon on
the compressor illuminates to indicate the ventilator
is charging. Ensure the compressor is powered on.
If the battery does not recharge, contact your service
representative.
Ensure the ventilator is properly docked and locked into the
compressor and the locked icon is illuminated to indicate
the ventilator is charging.
Ensure the compressor is powered on.
If the battery does not recharge, contact your service
representative.
Values displayed on the screens are for illustrative purposes only.
HIGH-PRIORITY ALARMS (CONTINUED)
NOTIFICATION
CAUSE
CHECKS AND POSSIBLE RESOLUTION IN EACH CONFIGURATION
STAND-ALONE CONFIGURATION
EXTENDED RANGE CONFIGURATION
STATIONARY CONFIGURATION
Very Low Battery
10:11a, Fri
Jun 12, 2015
12
Breath/min
Air LPM
PIP cmH
2
O
3.9
1.1
Very Low Battery
Ventilator battery capacity drops
below 15%
Connect the ventilator to the ventilator battery
charger and an AC power source to recharge the
battery.
Or dock the ventilator back into the compressor
to recharge the battery. Ensure that the locked
icon on the compressor illuminates to indicate the
ventilator is charging. Ensure the compressor is
powered on.
If the battery does not recharge, place the patient
on an alternate means of ventilation (if necessary)
and contact your service representative.
Connect the ventilator to the ventilator battery
charger and an AC power source to recharge the
battery.
Or dock the ventilator back into the compressor to
recharge the battery. Ensure that the locked icon on
the compressor illuminates to indicate the ventilator
is charging. Ensure the compressor is powered on.
If the battery does not recharge, place the patient on
an alternate means of ventilation (if necessary) and
contact your service representative.
Ensure the ventilator is properly docked and locked into the
compressor and the locked icon is illuminated to indicate the
ventilator is charging.
Ensure the compressor is powered on.
If the battery does not recharge, place the patient on an
alternate means of ventilation (if necessary) and contact your
service representative.