ALARMS, ALERTS, AND TROUBLESHOOTING
7
ALARMS, ALERTS, AND TROUBLESHOOTING
7
ADDITIONAL TROUBLESHOOTING SITUATIONS (CONTINUED)
NOTIFICATION
CAUSE
CHECKS AND POSSIBLE RESOLUTION IN EACH CONFIGURATION
STAND-ALONE CONFIGURATION
EXTENDED RANGE CONFIGURATION
STATIONARY CONFIGURATION
Negative PIP value is
displayed
Low PIP
When using the Life2000® Ventilator interfaced
with the UCC in conjunction with a full-face mask;
when the system is disconnected, the Low PIP
alarm will be initiated and a negative value PIP
may be displayed.
When using the Life2000® Ventilator interfaced
with the UCC in conjunction with a full-face mask;
when the system is disconnected, the Low PIP
alarm will be initiated and a negative value PIP
may be displayed.
When using the Life2000® Ventilator interfaced with the
UCC in conjunction with a full-face mask; when the system
is disconnected, the Low PIP alarm will be initiated and a
negative value PIP may be displayed.
"Incompatible Settings"
message is displayed
The PEEP or Volume setting is
incompatible to maintain clinical
settings.
Touch
OK
and then re-evaluate and readjust the
Volume, PEEP, or I-Time setting in the Ventilation
Settings screen.
Touch
OK
and then re-evaluate and readjust the
Volume, PEEP, or I-Time setting in the Ventilation
Settings screen.
Touch
OK
and then re-evaluate and readjust the Volume,
PEEP, or I-Time setting in the Ventilation Settings screen.
"This Prescription
Setting is not active"
message is displayed
The selected Prescription Setting
button is not active.
Touch
OK
and choose another active Prescription
Setting button to change the currently-ventilating
prescription.
Touch
OK
and choose another active Prescription
Setting button to change the currently-ventilating
prescription.
Touch
OK
and choose another active Prescription Setting
button to change the currently-ventilating prescription.
"Connect Oxygen
Source" or "Disconnect
Oxygen Source"
message is displayed
A reminder to verify that the correct
source gas is connected
Verify that the correct source gas is connected,
then press
OK
to begin ventilating.
Verify that the correct source gas is connected,
then press
OK
to begin ventilating.
Verify that the correct source gas is connected, then press
OK
to begin ventilating.