Example Setup of PEM-100 Controller for use with the CLI
PEM 100 User Manual
51
PEM-100 ‘Command Line Interpreter (CLI)’ Command Descriptions
Action/terminal
characters typed
Command Description
Expected Terminal
Display
PEM-100
Remote Mode
Display
Remark
Enter key pressed
Enter
*(ready indicator)
No change
Incorrect syntax or non-
accepted values
? (error indicator)
No change
Is always followed by a ‘*’ character on the
next line.
F and 1F
Display 1f frequency on the
terminal display and on the
Remote control mode display
Displays 1f freq in 6 digits.
039988 for example.
Displays 1f freq
in kHz
Correctly entered command followed by the 1f
frequency, then a ‘*’ character on the next line.
Allow a few seconds for frequency stability.
2F
Display 2f frequency on the
terminal display and on the
Remote Control mode display.
Displays 2f freq in 6 digits.
079976 for example.
Displays 2f freq
in kHz
Correctly entered command followed by the 2f
frequency, then a ‘*’ character on the next line.
Allow a few seconds for frequency stability.
F?
Displays the 1f or 2f frequency on
the terminal display if the 1F or 2F
command is active.
Displays freq in 6 digits.
039988 for example.
No change
Correctly entered command followed by the
currently selected frequency, then a ‘*’
character on the next line. Allow a few
seconds for frequency stability.
I:n, n=0 or 1.
Inhibit retardation. 0 = normal
operation, 1 = inhibit
None
No change
Correctly entered command followed by a ‘*’
character on the next line.
R:nnnn, nnnn – 0000 to
1000
Sets retardation in wave units. Full
wave = 1000, ¼ wave = 0250.
None
No change
Correctly entered command followed by a ‘*’
character on the next line.
W:nnnnnn, nnnnnn =
00000.0 to 19999.9.
Sets retardation in nm units.
None
No change
Correctly entered command followed by a ‘*’
character on the next line.
R
Reads the current PEM-100
retardation setting in wave units.
Result of retardation query
is displayed as nnnn.
No change
Correctly entered command results in the nnnn
value of retardation being displayed and
followed with the ‘*’ character on the next line.
W
Reads the current PEM-100
wavelength setting in nm units.
Results of wavelength
query is displayed as
nnnnnn.
No change
Correctly entered command results in the
nnnnnn value of wavelength being displayed
and followed with the ‘*’ character on the next
line.
Table 8.2a PEM-100 Command Line Interpreter Command Descriptions
Summary of Contents for PEM-100
Page 1: ...PEM 100 PHOTOELASTIC MODULATOR USER MANUAL Hinds Instruments Inc P N 010 0000 021 UM Rev H ...
Page 4: ......
Page 8: ...Table Of Contents viii PEM 100 User Manual ...
Page 12: ...Table of Tables xii PEM 100 User Manual ...
Page 22: ...Unpacking the PEM 100 10 PEM 100 User Manual ...
Page 46: ...Chapter 5 PEM 100 Controller Display 34 PEM 100 User Manual ...
Page 52: ...Chapter 6 Operating the PEM 100 Controller 40 PEM 100 User Manual ...
Page 68: ...Chapter 9 Troubleshooting 56 PEM 100 User Manual ...
Page 74: ...Chapter 10 Maintenance 62 PEM 100 User Manual ...
Page 88: ...Appendix A Calibration 76 PEM 100 User Manual ...
Page 92: ...Appendix C Optical HeadSpecifications 80 PEM 100 User Manual ...
Page 100: ......