PEM200 User Manual
46
Cotton Swab Method
Using a 100% cotton swab moistened with solvent, wipe the surface gently while
rotating the swab slowly. This action ensures that the contaminants are lifted away from
the surface of the optic.
Figure 9.1 Direction of cotton swab wiping motion
Lens Tissue Method
In order to clean the optics using the lens tissue method, the cover to the optics must
first be removed.
When applying the solvent, avoid exposing the junction between the optical element
and the transducer.
Figure 9.2 Junction between Optical Element and Transducer
1. Take a single piece of lint-free lens tissue (do not use Kimwipes) and place it on the
surface to be cleaned, so it covers the entire surface.
2. Using the recommended cleaning agent, soak the lens tissue until the entire surface
is damp (not wet).
3. Grasp two free corners of the lens tissue and drag the tissue across the surface
away from the junction in a smooth, even manner so the alcohol dries immediately
as the rear edge of the tissue moves over the optical surface.
Direction of
wiping motion
Transducer
Optical Element
Junction