range of the
instrument.
EC02 Air circuit leakage, which may be a
result of leakage of valve or air circuit.
Return to the factory
EC03 Wrong air pressure, which may be a
result of normal open failure of valve.
Return to the factory
EC04 Weak signal, which may be a result of
too weak pulse of the measured object or
loosen cuff.
Confirm that the
measurement part is
in good contact with
the cuff and whether
the arm
circumference is
within the measuring
range
EC05 Out-of-range, which may be a result of
blood pressure of the measured object
exceeding measurement range
Remeasurement
EC06 Excessive motion, which may be a result
of motion artifacts or more interferences
contained in signals during measurement
Note to avoid
speaking and
disturbance during
the measurement
process and measure
again
EC07 Overpressure in measurement, namely
cuff pressure in adult mode exceeding
290mmHg
EC09 Timeout in measurement, namely
measurement time in adult mode
exceeding 120 seconds
remeasurement
EC10 Manual stop
1. Touch the
start/stop button
2. Detect whether the
button is insensitive
or stuck
EC11 System error
remeasurement
EC16 Overpressure protection, which is caused
by cuff pressure exceeding the set
maximum value (290)
EC17 Sleeve failure, motor failure
remeasurement
EC19 The arm posture is not correct and the
elbow switch is not pressed
Adjust the arm
posture, make sure
the elbow switch is
pressed
EC32 Communication failure, handshake
communication failure
remeasurement
EC35 Failed to start, no response after sending
measurement, unable to start
measurement
remeasurement