(5) Please place the printing paper after the alcohol has dried up.
(6) Lift up the edge of the paper, close the printer cover, and make sure that the
printing paper is exposed.
Note: When cleaning the print head, static electricity will cause damage to the print
head. Please note the static electricity.
9.4 Scheduled maintenance
To use the instrument correctly, perform regular inspection. Regular inspection
mainly comprises the following items:
Before power on
Items
Contents
Appearance
Is there any deformation or damage caused by falling,
etc.?
Are parts dirty, rusted or scratched
Is screen dirty or damaged
Is it wet?
Operation parts
Are switches and buttons damaged or crashed?
Measurement parts
Is the cuff installed correctly?
Is there any damage, obviously dirt or blood stains on
the cuff?
Power cable
Is the power cable properly connected or damaged or
provided with grounded 3P socket?