background image

3

Table of Content 

Chapter 1 Preface ..................................................................................................................5 

1.1 Brief Introduction to Ambulatory Blood Pressure Monitoring ...................................5 

Chapter 2 Safety Requirements ............................................................................................7 

2.1 Intended use .............................................................................................................7 

2.2 Contraindication ........................................................................................................8 

2.3 Warnings ...................................................................................................................8 

2.4 Notice ........................................................................................................................9 

2.4.1 Battery ............................................................................................................9 

2.4.2 Blood pressure measurement ........................................................................9 

2.4.4 Energy conservation and environment protection .........................................9 

2.4.5 Maintenance ...................................................................................................9 

Chapter 3 Product Introduction ........................................................................................... 10 

3.1 Product composition: ............................................................................................. 10 

3.2 Name and Function of Each Component .............................................................. 10 

3.2.1 Host Introduction ......................................................................................... 10 

3.2.2 OLED screen displays and icon implication ................................................. 11 

3.2.3 Symbols ........................................................................................................ 11 

Chapter 4 Product Installation and Use .............................................................................. 12 

4.1 Battery installation ................................................................................................. 12 

4.2 Wear the monitor ................................................................................................... 12 

4.2.1 Wear cuff correctly ...................................................................................... 12 

4.2.2 Start for single measurement ...................................................................... 12 

4.2.3 Ready for use .............................................................................................. 12 

4.3 Installing System (PC terminal) ............................................................................. 13 

4.3.1 Hardware requirements ................................................... 

错误!未定义书签。

 

4.3.2 Software requirements .................................................... 

错误!未定义书签。

 

4.3.3 Install the software .......................................................... 

错误!未定义书签。

 

4.4 Execute Ambulatory Blood Pressure Study (PC terminal) ........ 

错误!未定义书签。

 

4.4.1 Communication with Hingmed ABPM ............................. 

错误!未定义书签。

 

4.4.2 Connect monitor to computer .......................................... 

错误!未定义书签。

 

4.4.3 Continuous measurement programming ........................ 

错误!未定义书签。

 

4.4.4 Start monitoring ............................................................... 

错误!未定义书签。

 

4.4.5 Finish measuring ............................................................. 

错误!未定义书签。

 

4.4.6 Derive data ...................................................................... 

错误!未定义书签。

 

4.5 Install system (Smart mobile terminal) .................................................................. 13 

4.5.1 Hardware requirements: .............................................................................. 13 

4.5.2 Installation of software................................................................................. 13 

4.5.3 The Bluetooth matching and connection ............................................................ 13 

4.6 Execute Ambulatory Blood Pressure study (Smart mobile terminal). ................... 14 

4.6.1 Communicate with Household Ambulatory Blood Pressure Monitor .......... 14 

4.6.2 Program Ambulatory Blood Pressure Monitor software .............................. 14 

4.6.3 Start ambulatory blood pressure measurement .......................................... 14 

Summary of Contents for WBP-03D

Page 1: ...1 CE0120 Hingmed Wearable Ambulatory Blood Pressure Monitor User Manual Apply to WBP 03D for home use Shenzhen Hingmed Medical Instrument Co Ltd VER A 0 Date March 15 2015 ...

Page 2: ...518102 After sales service Company Shenzhen Hingmed Instruments Co Ltd Address 4th Floor Zhonghangfeixiang Building NO 371 Guangshen Road Bao an District Shenzhen People s Republic of China Tel 86 755 23720600 Fax 86 755 23720602 Post code 518102 European Representative Company Name Well Kang Limited Company Address The Black Church St Mary s Place Dublin 7 Ireland Tel 353 1 4433560 Fax 353 1 6864...

Page 3: ...rement 12 4 2 3 Ready for use 12 4 3 Installing System PC terminal 13 4 3 1 Hardware requirements 错误 未定义书签 4 3 2 Software requirements 错误 未定义书签 4 3 3 Install the software 错误 未定义书签 4 4 Execute Ambulatory Blood Pressure Study PC terminal 错误 未定义书签 4 4 1 Communication with Hingmed ABPM 错误 未定义书签 4 4 2 Connect monitor to computer 错误 未定义书签 4 4 3 Continuous measurement programming 错误 未定义书签 4 4 4 Start mon...

Page 4: ...d data derive 15 Chapter 5 Troubleshooting and Maintenance 15 5 1 Troubleshooting 15 5 2 Maintenance 16 5 2 1 Maintenance inspection and safety management 16 5 2 2 Host Maintenance 16 Chapter 6 Product specifications 18 Chapter 7 EMC information 19 Chapter 8 Warranty Card 23 ...

Page 5: ...n Accordingly patients could take short medium or long term antihypertensive drugs to control their blood pressure and avoid damage of target organs e g cardio brain and kidney Obviously Ambulatory blood pressure monitoring is superior to clinical blood pressure measurement and gradually becomes an important method to evaluate the treatment effect of ant hypertension gradually Compare to clinical ...

Page 6: ...d pressure In daytime patients are active and in nighttime patients are sleeping it is also influenced by external causes i e sleep wake cycle and internal cause i e biological clock Specific circadian rhythm variability is the average change rate of blood pressure during nighttime and daytime Although the circadian rhythm is common for normal people and primary hypertension patients some special ...

Page 7: ...hypertension Position hypertension is that the pressure is high at standing or sitting posture but normal at lying flat Orthostatic hypertension accounts for 4 2 in domestic hypertension patients and 10 in abroad report Orthostatic hypertension is characterized by that there is not hypertension in normal and just being found occasionally or in physical examination Diastolic pressure rising up and ...

Page 8: ...It may cause measurement error by using parts not included in the supply listing Make sure the pressure is suitable for the patient If any abnormity occurs in the monitoring process please stop measuring immediately and remove the cuff from the patient or shortly press start stop button to stop inflating If the cuff fails to deflate please remove it properly and safely If this device is dampened b...

Page 9: ...n cannot be charged or its discharge speed is fast 2 4 2 Blood pressure measurement Patient with anticoagulant or patient with coagulation disorder may extravasate on the position wrapping cuff in the process of measuring blood pressure even the cuff is worn correctly In fact no matter what type of the monitor is such patients would extravasate during the measuring process If cuff fails to inflate...

Page 10: ...ories User Manual Quality Certification warranty card Hingmed ABPM Packing List Accessories Items Name Qty Hingmed Wearable ABPM Monitor 1 Portable box for monitor 1 USB data wire 1 Adult cuff 26 36 cm 1 User Manual 1 Quality Certification 1 The product packaging should contain above items any item misses please contact Hingmed or distributors immediately Optional accessories Number Accessories Na...

Page 11: ...ndicated after measuring Heart rate unit times per minute Indicate current time when not measuring after powering on indicate battery capacity 3 2 3 Symbols Symbol Description Symbol Description CF type applied part Series number Manufacturer Attention refer to accompanying documents Keep away from rain This way up Fragile Complying with WEEE standard Stacking limit by number Temperature limits Re...

Page 12: ...rement Press start stop button to start measuring the middle button it only measures for one time after measuring user can read the data on OLED screen also user can click forward button or backward button to read history data For programming measurement please refer to sections 4 3 4 4 of this user manual 4 2 3 Ready for use For ambulatory blood pressure study a good preparation of patient is imp...

Page 13: ...APP platform download and install Hingmed blood pressure software For Apple IOS system handhold terminal Users log on App Store download and install Hingmed blood pressure software The icon appears on the screen of handhold terminal means a successful software installation 4 3 3 The Bluetooth matching and connection For Android phone enter Setting Bluetooth at the time keep the Bluetooth of monito...

Page 14: ... successive measurements The options may be 15 30 45 60 90 and 120 min 3 Maximum inflation pressure Maximum inflation pressure is the maximum one in a measurement Usually the range is set from 200mmHg to 280mmHg Possible maximum systolic pressure plus 40mmHg is recommended The actual inflation pressure setting should not reach up to the maximum pressure which may cause user discomfort You can set ...

Page 15: ...ly until programming it from smart mobile phone Chapter 5 Troubleshooting and Maintenance 5 1 Troubleshooting Error Code Description Resolution EC01 Cuff is too loose maybe loose winding or disconnected cuff Retighten cuff or keep proper arm position when inflation EC02 Air leakage maybe valve leakage or gas path leakage Tighten metal connector and check the cuff If leakage is not still resolved c...

Page 16: ...urement result is not available Press start stop button to measure again EC37 Overtime beyond 180s remeasure EC48 Memory is full can no longer measure Upload and empty the data If the errors can be revised please call distributor or after sale department of Hingmed for help 5 2 Maintenance 5 2 1 Maintenance inspection and safety management Visually inspect the broken or damages on enclosure and cu...

Page 17: ...ter drying put the bladder into the cuff then install the monitor onto the cuff See the following steps to take down the cuff a Dismantle NO Picture Description NO Picture Description 1 Hold the positions shown in the picture and give force by following the red arrows 2 Pull out the top two fixing clips 3 Then move down along the unit hold the positions shown in the picture and give force by follo...

Page 18: ...s low pressure mode Accuracy Static Accuracy 3 mmHg Pulse rate 3BPM Pressure testing American Freescale pressure sensor Working power supply Built in Li ion battery Data memory Flash memory stores up to 300 readings Calibration frequency At least one time two years Security system The rated range of cuff pressure 0 290mmHg Sampling period Multiple independent programming cycle Size About 119 52 21...

Page 19: ... should be observed to verify normal operation in the configuration in which it will be used Guidance and manufacture s declaration electromagnetic emission The WBP 03D is intended for use in the electromagnetic environment specified below The customer of the user of the WBP 03D should assure that it is used in such an environment Emission test Compliance Electromagnetic environment guidance RF em...

Page 20: ...ent Surge IEC 61000 4 5 1 kV line s to line s 2 kV line s to earth Not applicable Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 sec Not applica...

Page 21: ...commended separation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol Radiated RF IEC 61000 4 3 3 V m 80 MHz to 2 5 GHz 3 V m NOTE 1 At 80 MHz and 800 MHz the higher frequency range a...

Page 22: ...mitters and the WBP 03D as recommended below according to the maximum output power of the communications equipment Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 KHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output po...

Page 23: ... original purchaser should notify Hingmed of the suspected defect the monitor should be carefully packaged and be prepaid shipped to Shenzhen Hingmed Medical Instrument Co Ltd Address 4th Floor Zhonghangfeixiang Building NO 371 Guangshen Road Bao an District Shenzhen People s Republic of China Tel 86 755 23720600 Fax 86 755 23720602 Postal code 518102 Email info Hingmed com The monitor will be rep...

Reviews: