9
●
Inflate the cuff on the side of a mastectomy.
●
Do not inflate the cuff on the same limb which other monitoring ME equipment is
applied around simultaneously, because this could cause temporary loss of function of
those simultaneously-used monitoring ME equipment.
●
Please check that operation of the device does not result in prolonged impairment of
patient blood circulation.
●
The device cannot be used with HF surgical equipment at the same time.
●
Keep out of the reach of children after use.
2.4 Notice
2.4.1 Battery
●
The monitor uses built-in battery, do not replace it without authorization.
●
Connect monitor host to computer USB port through USB data wire for recharging,
alternatively, connect host to USB port of adapter, and then connect adapter to power
socket for recharging. The adapter should be with 3C compulsory certification or other
compulsory certifications (such as FCC or CE). The specifications of adapter should be:
input: 100-240V
, 50/60HZ; output: 5V,
1A.
●
Keep the monitor away from high-temperature places, and avoid direct-sunlight in
summer, also, the environmental temperature shall not be over 60 Celsius degree.
●
For security, consult after-sale supporter quickly, if the Li-ion cannot be charged or its
discharge speed is fast.
2.4.2 Blood pressure measurement
●
Patient with anticoagulant or patient with coagulation disorder may extravasate on the
position wrapping cuff, in the process of measuring blood pressure even the cuff is
worn correctly. In fact, no matter what type of the monitor is, such patients would
extravasate during the measuring process.
●
If cuff fails to inflate in 150 seconds, guide patients to remove the cuff manually,
excessive inflation may block patient's blood flow that makes patient uncomfortable.
●
Keep still during measuring, especially do not move their cuffed arms, and it is better to
keep quiet, and make the cuff as the same horizontal position as the heart.
●
Operating or storing the monitor beyond the specified environmental conditions in
Chapter 6 would cause damage.
2.4.4 Energy conservation and environment protection
●
Please power the monitor off in time after using.
●
Please properly handle the scrapped batteries, cuff, data wires and main units in
compliance with local environmental regulations.
2.4.5 Maintenance
●
Please use the accessories supplied by Hingmed, otherwise it may cause measuring
errors.
●
Maintenance should be conducted only by trained personnel or personnel authorized
by Hingmed.
●
Displaying method of faults: corresponding codes will display on the screen in
malfunction state, see 5.1 troubleshooting for details.