37
4.2 Accuracy
Accuracy Table
(Accuracy guaranteed for one year at 23°C±5°C (73°F±9°F),
80%RH or less.)
4.2 Accuracy
(rdg.: displayed value, dgt.: resolution)
Voltage Measurement
Range
[V]
Accuracy
±(rdg.)±(dgt.)
Input Impedance
(Frequency range)
mV
V
420.0 m
4.200
42.00
420.0
1000
±1.0%±4
±1.0%±4
±0.5%±4
±0.7%±4
±0.7%±4
Approx.10 M
Ω
Approx.11 M
Ω
Approx.10 M
Ω
Approx.10 M
Ω
Approx.10 M
Ω
mV
V
420.0 m
4.200
42.00
420.0
1000
±2.0%±4
±2.0%±4
±1.2%±4
±1.5%±4
±1.5%±4
(50
to
500 Hz)
Approx.10 M
Ω
Approx.11 M
Ω
Approx.10 M
Ω
Approx.10 M
Ω
Approx.10 M
Ω
Overload protection (for one minute):
1000 VDC,1000 Vrms(sin) or 10
7
V
•
Hz
Summary of Contents for 3255-50 HiTESTER
Page 1: ...INSTRUCTION MANUAL DIGITAL HiTESTER 3255 50 ...
Page 2: ......
Page 16: ...Usage Notes 12 ...
Page 20: ...1 3 Parts Names and Functions 16 ...
Page 36: ...3 4 Auto Power Save Function 32 ...
Page 48: ...5 3 Service 44 ...
Page 49: ......
Page 50: ......