21
2.3 Current Measurement
mA Measurement (40 mA/400 mA)
1.
Move the function selector to the
mA
position.
2.
Select DC (
) or AC (
) using
the
REC•MEMO
button.
3.
Connect the red test lead to ter-
minal
μ
A.mA
, and the black test
lead to terminal
COM
.
4.
Connect the test leads to the
measurement object, and read
the indicated value.
REC•MEMO
Black
Red
A Measurement (10 A range)
Note that the product may be damaged if current
exceeding the selected measurement range is
applied for a long time (for the 10 A range, continu-
ous current must be limited to 7A, or to less than
one minute if over 7A)
REC•MEMO
Black
Red
1.
Move the function selector to the
10A
position.
2.
Select DC (
) or AC (
) using
the
REC•MEMO
button.
3.
Connect the red test lead to termi-
nal
A
, and the black test lead to
terminal
COM
.
4.
Connect the test leads to the mea-
surement object, and read the
indicated value.
The Auto Range function is not
provided.