39
3.6 Dynamic Recording Function
3.6 Dynamic Recording Function
Functions
V/
V/
/40
A/
A/mA/10A
Description
This function allows the maximum (MAX), mini-
mum (MIN), average (AVG), and currently mea-
sured values to be displayed selectively. It is
useful for measuring any discrepancy over an
extended period.
Appearance of O.F: The measurement range exceeds the
specified level.
Before using the Recording function, use
the manual range to set the range as-
sumed as the maximum level.
Canceling
Press
REC•MEMO
again.
1.
Select the desired function
and connect the test leads to
the measurement product.
2.
Connect the test leads to the
measurement object.
3.
Press the
REC•MEMO
but-
ton for at least 1 second.
("MAX MIN AVG" lights up)
4.
Select the item using the
H.AUTO/REL
button.
Current
value
Maximum value
Minimum value
Average value
REC•MEMO
Press
for at least
1 second.
H.AUTO/REL
• The maximum, minimum, and aver-
age measured after the activation of
the Recording function are automati-
cally recorded (recording interval:
approx. 0.4 seconds).
• When the maximum and/or minimum
is updated, a beeping sound is gen-
erated.
• The simple average is displayed
within 30 min from the start of mea-
surement and the moving average is
displayed after the elapse of 30 min.
• While the Recording function is in
operation, the Auto Power Save
function is disabled (APS is turned
off). The range is fixed.