2.7 Diode Check
30
2.7 Diode Check
• Never apply voltage to test leads when
the Diode Check function is selected.
Doing so may damage the product and
result in personal injury.
• To avoid electrical accidents, remove
power from the circuit before measuring.
Black
Red
REC•MEMO
Cathode Anode
Black
Red
1.
Move the function selector to the
position
2.
Select Diode (
) using the
REC•MEMO
button (
lights up).
3.
Connect the red test lead to ter-
minal
V.
, and the black test lead
to terminal
COM
.
4.
Connect the test leads to the
measurement object.
The display shows forward volt-
age (0.4 to 0.7 V) for a normal
diode.
Appearance of O.F:
The diode is invertedly connected or
broken.
Appearance at approxi-
mately 0 V:
The diode is short-circuited.