Chapter 2Set Up
______________________________________________________________
WARNING
The maximum input voltage is 45 V DC. Attempting
to measure voltage in excess of the maximum input
could destroy the product and result in personal
injure or death.
9629 CONNECTION CABLE
Black
Red
For connection between the 9629
CONNECTION CABLE and the pulse-generating
device, refer to the operation manual of the
pulse-generating device.
2.4 Connecting Sensor Cable
Connecting the CONNECTION CABLE
The 3639-20 PULSE LOGGER is capable of measuring one
channel of pulse signals with the 9629 CONNECTION
CABLE connected. To ensure correct measurement, insert
the CONNECTION CABLE as far as it will go.
The 9629 CONNECTION CABLE can be cut to any
desired length for use. If the cable was cut, use it after
stripping the covering with a wire stripper or another
appropriate tool.
To input a voltage to the 3639-20, connect the red
conductor to the positive (+) side of the pulse-generating
device, and connect the black conductor to the negative (-)
side. If the conductors are inversely connected,
measurement cannot be conducted.