13
_____________________________________________________________________
Chapter 3Settings
______________________________________________________________
5. Start control
Set specific recording date and time using 3911-20 to
engage time scheduled start. When time scheduled start is
engaged,
appears in display.
When the Power Saving function is set to OFF before
measurement starts, pressing the INTERVAL button turns
the Power Saving function ON/OFF. When the
measurement-value display is on, the Power Saving
function is set to OFF. When the measurement-value
display is off, the Power Saving function is set to ON. For
an extended measurement operation, set the Power Saving
function to ON (measurement-value display turned off).
6. Recording method setting
Set recording method using 3911-20.
Choose either one time or endless recording method.
Default setting is one time.
One time: Ends recording when data reaches 32000.
Endless : Overwrites previously recorded data when data
exceeds 32000.
7. Comments
The settings Comments is available only when the three
products of the PC, the 3911-20, and the 3639-20 are
interconnected.
Set comments entered by personal computer to logger using
3911-20. When sorting collected recording data, comments
are helpful.