16
_____________________________________________________________________
Chapter 3Settings
______________________________________________________________
3.2 Manual Setting
3639-20 PULSE LOGGER stand-alone manual operation
settings are shown below.
(1) Interval setting
Press INTERVAL button to switch measurement value
display to interval setting display. (
INTVL
appears.)
Press SELECT button to designate interval.
Press INTERVAL button to complete setting.
(2) Starting and ending recording
Press REC/STOP button for 1 second to clear last
recorded data and start recording. (
REC
appears.)
Press REC/STOP button for 1 second to stop recording.
When memory is full, recording automatically stops
when recording method: one time is selected.
Recording cannot be started if the battery power is low
(
indication). Moreover, if the battery power
becomes too low during recording (
indication),
recording will be interrupted.
When the measurement value display is on, the
displayed values are updated at every recording interval.
At the outset of a measurement session, 0 will be
displayed.