3.4 Recording Function
43
1
2
3
4
5
6
7
8
9
10
11
A
dditiona
l Func
tions
This function records the maximum, minimum and aver-
age value of the input signal starting from the time when
the recording function was started.
Input signal current value
:
MAX MIN AVG
Input signal maximum value :
MAX
Input signal minimum value
:
MIN
Input signal average value
:
AVG
1.
To turn on the recording function, hold down the
HOLD
key while measurement is in progress.
• The range is locked in the peak hold function. Press
the
RANGE
key to select the range. Changing the
range restarts the function.
• The sub display shows the time from the start of the
peak hold function until the maximum (minimum)
value was updated. The maximum elapsed time that
can be displayed is 99999 seconds; if this time is
exceeded, "
OL
" is displayed.
• The auto power save function is disabled automati-
cally when you use the peak hold function.
• Only DC function accuracy is specified for the peak
hold function.
3.4 Recording Function
Elapsed time
or update time