2.1 Pre-Operation Inspection
16
Operation Check
If the operation check reveals any abnormalities,
stop the check immediately and do not use the
instrument.
Required equipment:
• Model 3804-50 (this instrument)
• Model 3851-10 TEST LEAD
• AC power receptacle
(100 V AC 50 Hz/60Hz commercial power supply)
1.
Set the function switch to "
Ω
".
2.
Press the
SHIFT
key to conduct the continuity
check. (
lights.)
To avoid electrocution, do not touch the ter-
minals.
For safety reasons, when taking measurements,
only use the test lead provided with the instrument.
In order to protect the tips
of the test leads, the test
leads are capped when
the unit is shipped from
the factory. Be sure to
remove the caps before
using the test leads.