2.1 Pre-Operation Inspection
18
7.
(Confirming blown fuse for mA, mA terminal)
Set the function switch to "Diode Check (
)".
Connect the red test lead to the µA, mA terminal,
and the black test lead to the V terminal. Short the
tips of the red and black test leads by touching
them together.
8.
(Confirming blown fuse for A terminal)
Connect the red test lead to the A terminal, and
the black test lead to the V terminal. Short the tips
of the red and black test leads by touching them
together.
This procedure only partially confirms the opera-
tion of this instrument. Periodic calibration is nec-
essary in order to ensure that this instrument
operates according to its product specifications.
•
Approximately 1.6 V displayed
→
OK
•
OL
displayed
→
NO
Fuse may be blown
.
Please replace fuse.
•
Buzzer sounds
→
OK
•
OL
displayed
→
NO
Fuse may be blown.
Please replace fuse.