5.5 Value Comparison Details
______________________________________________________________________________________________
Increases in number
Move the cursor to right
Decreases in number
Edit value of specified level
Move the cursor to left
Confirms the value
NOTE
NOTE
2. Setting Specified level (Level)
Set the specified level to use as a reference for period, frequency, pulse
width, duty ratio, specified level time, and pulse count.
Setting Procedure
1. Move the cursor to "
Level
"
(Note)
, then
press the
ENTER
key.
2
.
Use the
CURSOR
key to change the
numeric value, then press the
ENTER
key
to confirm the change.
(Note)This can only be selected when the type
of calculation uses the specified level.
When using waveform calculations (See 5.6),
make the setting using LSB (minimum
resolution).
3. Setting Slope
Select the slope direction to serve as the measurement reference for period,
frequency, pulse width, specified level timing and pulse count.
Setting Procedure
1. Move the cursor to "
Slope
"
(Note)
, then
press the
ENTER
key.
2
.
Use the
CURSOR
key to select the slope,
then press the
ENTER
key.
UP : Up-slope
DOWN: Down-slope
(Note) This selection can be made when the
calculation type involves the slope direction.