8.10 Making Detailed Settings for Input Modules
148
See:
Opening the
[Each Ch]
sheet, Making a Channel Selection (
Mode
There is no need to change the setting since it is set when the clamp sensor is
automatically recognized.
Select
Range
Change the measurement of the 8971 Current Unit between DC and RMS.
Select
8.10.6 Settings for the 8971 Current Unit
Selections
Description
20A/2V
Set when a 9272-10 (20 A range) or 9277 Clamp Sensor is con-
nected. (default setting)
200A/2V
Set when a 9272-10 (200 A range), 9277, or CT6863 Clamp Sen-
sor is connected.
50A/2V
Set when a CT6862 Clamp Sensor is connected.
500A/2V
Set when a 9279 or 9209 Clamp Sensor is connected.
Selections
Description
DC
Current measurement (default setting)
RMS
RMS measurement