9.2 Setting the Trigger Mode
153
9
Chapt
er
9
T
rigger
Set
tings
Set whether to continue to accept triggers after measuring.
If all trigger sources are disabled (Off, with no trigger setting), measurement starts immediately
(free-running).
Description
Available selections depend on the function.
To Stop Measuring:
Press the
STOP
key.
Press once: recording stops at the end of the specified recording length.
Press twice: recording stops immediately.
When the trigger mode is set to
[Repeat]
During the processing interval from the end of recording until entering the next
trigger standby condition (Auto Save, Auto Print, waveform display processing,
numerical calculation), triggering will not be accepted.
9.2
Setting the Trigger Mode
Procedure
To open the screen: Press the
DISP
key
→
Waveform screen
1
Move the cursor to the
[Trigger]
item.
2
Select the trigger mode.
Select
Single
Only one trigger is recognized. After pressing the
START
key, once a trigger is applied, a waveform is recorded for
the specified recording length, and measurement then
stops.
Repeat
Triggers are accepted continuously.
When no trigger is applied, the instrument enters the Trig-
ger Wait state. Press the
STOP
key to stop measuring.
(See below)
Auto
Triggers are accepted continuously.
If no trigger is applied within about one second, a waveform
of the specified recording length is automatically recorded.
Press the
STOP
key to stop measuring.
Trigger Mode
Function
Single
O
O
(default setting)
Repeat
O
O
Auto
O
(default setting)
×
Summary of Contents for 8847
Page 1: ...MEMORY HiCORDER Instruction Manual 8847 December 2010 Revised edition 4 8847A981 04 10 12H ...
Page 2: ......
Page 18: ...Operating Precautions 8 ...
Page 26: ...1 4 Basic Key Operations 16 ...
Page 74: ...4 4 Waveform Observation 64 ...
Page 98: ...5 5 Managing Files 88 ...
Page 110: ...6 6 Miscellaneous Printing Functions 100 ...
Page 126: ...7 8 Seeing Block Waveforms 116 ...
Page 160: ...8 10 Making Detailed Settings for Input Modules 150 ...
Page 198: ...10 6 Numerical Calculation Type and Description 188 ...
Page 296: ...15 7 Controlling the Instrument with Command Communications LAN USB 286 ...
Page 304: ...16 2 External I O 294 ...
Page 362: ...Appendix 5 FFT Definitions A28 ...
Page 368: ...Index Index 6 ...
Page 369: ......
Page 370: ......
Page 371: ......