9.8 Pre-Trigger Settings
167
9
Chapt
er
9
T
rigger
Set
tings
Description
About pre-triggering and the recording period (recording length)
Difference between
[Pre-Trigger wait]
and
[Trigger wait]
When measurement is started, the specified pre-trigger length is recorded. This
period is indicated as the
[Pre-Trigger wait]
.
After the specified pre-trigger length has been recorded, the period indicated as
[Trigger wait]
continues until a trigger occurs.
See:
"Measurement and Internal Operations" (
Trigger Point
95%
95% of the recording length is
recorded before the trigger
point
50%
50% of the recording length is
recorded before and 50% after
the trigger point
-95% 95% of the recording length is
recorded after the trigger point
T
Pre-Trigger setting examples
• Trigger events during the specified pre-trigger recording period are ignored. To
enable recognition of such triggers, set Trigger Priority to
[On]
.
See:
Summary of Contents for 8847
Page 1: ...MEMORY HiCORDER Instruction Manual 8847 December 2010 Revised edition 4 8847A981 04 10 12H ...
Page 2: ......
Page 18: ...Operating Precautions 8 ...
Page 26: ...1 4 Basic Key Operations 16 ...
Page 74: ...4 4 Waveform Observation 64 ...
Page 98: ...5 5 Managing Files 88 ...
Page 110: ...6 6 Miscellaneous Printing Functions 100 ...
Page 126: ...7 8 Seeing Block Waveforms 116 ...
Page 160: ...8 10 Making Detailed Settings for Input Modules 150 ...
Page 198: ...10 6 Numerical Calculation Type and Description 188 ...
Page 296: ...15 7 Controlling the Instrument with Command Communications LAN USB 286 ...
Page 304: ...16 2 External I O 294 ...
Page 362: ...Appendix 5 FFT Definitions A28 ...
Page 368: ...Index Index 6 ...
Page 369: ......
Page 370: ......
Page 371: ......