9.10 Setting Combining Logic (AND/OR) for Multiple Trigger Sources
170
Analog, logic, external and timer trigger criteria can be combined by AND/OR logic to define com-
plex trigger criteria.
Description
When the trigger combining logic (Source (AND/OR)) is set to
[AND]
If trigger criteria are already met when you press the
START
key, no triggering
occurs. Triggering occurs only after all trigger sources have ceased to meet the
criteria at once, and are subsequently met again.
Setting Example:
To apply a trigger when the upslope (
↑
) of the waveform crosses zero volts
Triggering occurs as follows in the AND and OR cases.
9.10 Setting Combining Logic (AND/OR) for
Multiple Trigger Sources
Procedure
To open the screen: Press the
DISP
key
→
Waveform screen
→
Press the
TRIG.SET
key
→
Trigger settings window
(Memory Function case)
Move the cursor to the
[Trig Src]
item.
Select
OR
Triggering occurs when any one of the specified trigger
source criteria is met. (default setting)
AND
Triggering occurs only when all of the specified trigger
source criteria are met.
Channel
Trigger
Trigger Level
Slope
Filter
Ch1, Ch2
Level
0.00 V
↑
Off
[AND]
Ch2
Ch1
One waveform has crossed
above 0V as the other
crosses on the upslope
[OR]
Either waveform crosses 0V
on the upslope
0 V
0 V
T
T
T
T
Ch2
Ch1
Start Measurement
Trigger criteria have been met,
but no trigger occurs.
Trigger criteria have been met,
but no trigger occurs.
If both
[Start]
and
[Stop]
trigger timing criteria are combined, the simultaneous trigger
sources are logically ANDed.
Summary of Contents for 8847
Page 1: ...MEMORY HiCORDER Instruction Manual 8847 December 2010 Revised edition 4 8847A981 04 10 12H ...
Page 2: ......
Page 18: ...Operating Precautions 8 ...
Page 26: ...1 4 Basic Key Operations 16 ...
Page 74: ...4 4 Waveform Observation 64 ...
Page 98: ...5 5 Managing Files 88 ...
Page 110: ...6 6 Miscellaneous Printing Functions 100 ...
Page 126: ...7 8 Seeing Block Waveforms 116 ...
Page 160: ...8 10 Making Detailed Settings for Input Modules 150 ...
Page 198: ...10 6 Numerical Calculation Type and Description 188 ...
Page 296: ...15 7 Controlling the Instrument with Command Communications LAN USB 286 ...
Page 304: ...16 2 External I O 294 ...
Page 362: ...Appendix 5 FFT Definitions A28 ...
Page 368: ...Index Index 6 ...
Page 369: ......
Page 370: ......
Page 371: ......