18.1 Trouble Shooting
313
Chapte
r 18
Main
tenanc
e and
Ser
vi
ce
14
18
If the cause is unknown
Try performing a system reset. All settings are returned to their factory defaults.
See:
"18.2 Initializing the Instrument" (
Cannot Print, or Printing Malfunctions
Symptom
Check Item, or Cause
Remedy and Reference
Nothing prints on the
paper.
Is the paper reversed (back to front)?
Verify that the recording paper is loaded cor-
rectly.
"2.4 Loading Recording Paper" (
Printout is too light.
• Is the specified recording paper being
used?
• Is the print density setting correct?
• Is the print head dirty?
Try changing the print density setting.
"6.5 Making Printer Settings" (
Clean the print head.
Recording traces are too
wide.
The input signal may have a ripple com-
ponent.
Enable the filter in the input module settings.
Recording traces are dou-
bled.
Is the waveform printing density set to
[Light]
?
In this case, printed dots are spaced out
in the vertical direction. Therefore, a
slightly varying waveform prints some-
times as one line, and sometimes as two.
Change to waveform printing density setting
to something other than
[Light]
. (
[Printer]
sheet)
"6.5 Making Printer Settings" (
If Saving is Not Possible
Symptom
Check Item, or Cause
Remedy and Reference
If saving is not possible to
the storage media.
• Is your CF card Hioki certified?
• Is the storage media inserted properly?
• Is the storage media formatted?
• Is the remaining capacity of the storage
media too low?
• Has the number of files in the folder
reached the limit?
"2.3 Recording Media Preparation"
(
Cannot use USB memory
stick
Have settings for using USB memory stick
been made?
Others
Symptom
Check Item, or Cause
Remedy and Reference
Cannot use USB commu-
nications
Have USB communication settings been
made?
"15.6.1 Making USB Settings at the
Instrument" (
Cannot install USB driver
• Is the
[Interface]
item set to
[USB]
?
• Set
[Interface]
item to
[USB]
.
"15.6.1 Making USB Settings at the
Instrument" (
• Has driver installation failed?
• Access the
[device manager]
on the
PC and delete
[Other Devices]
.
Then reconnect the USB cable.
The response of remote
control is very slow.
Check the java settings.
About the java settings, see the applica-
tion disk supplied with the instrument.
Summary of Contents for 8847
Page 1: ...MEMORY HiCORDER Instruction Manual 8847 December 2010 Revised edition 4 8847A981 04 10 12H ...
Page 2: ......
Page 18: ...Operating Precautions 8 ...
Page 26: ...1 4 Basic Key Operations 16 ...
Page 74: ...4 4 Waveform Observation 64 ...
Page 98: ...5 5 Managing Files 88 ...
Page 110: ...6 6 Miscellaneous Printing Functions 100 ...
Page 126: ...7 8 Seeing Block Waveforms 116 ...
Page 160: ...8 10 Making Detailed Settings for Input Modules 150 ...
Page 198: ...10 6 Numerical Calculation Type and Description 188 ...
Page 296: ...15 7 Controlling the Instrument with Command Communications LAN USB 286 ...
Page 304: ...16 2 External I O 294 ...
Page 362: ...Appendix 5 FFT Definitions A28 ...
Page 368: ...Index Index 6 ...
Page 369: ......
Page 370: ......