18.3 Error Messages
316
27
When executing, select "No" as the
save property.
As other dialogs are displayed, selective sav-
ing cannot be carried out. Either set the selec-
tive saving during execution to [No] or close
the dialog and re-save.
29
The record length was restricted.
-
-
30
Auto-ranging failed.
Check the input signal.
"3.7 Measurement With Au-
tomatic Range Setting (Auto-
Ranging Function)" (
31
A/B cursor positions invalid.
The A/B cursors overlap. Check the cursor po-
sitions.
"7.2 Specifying a Waveform
Range (A/B Cursor)" (
32
Zero-adjustment needed.
Perform zero-adjustment.
"2.7 Adjusting the Zero Po-
sition (Zero-Adjust)" (
33
Disabled key.
Close the dialog.
-
34
Invalid key pressed (Overlay)
The key operation is prohibited because Over-
lay is enabled.
"8.3 Displaying New Wave-
forms Over Past Waveforms
(Overlay)" (
36
No trigger has been set.
Set trigger criteria.
"Chapter 9 Trigger Set-
tings" (
38
Use of logic channels has reduced ana-
log waveform resolution from 16 to 12
bits.
Using channels LA - LD reduces resolution of
Ch1 - Ch4 to 12 bits.
"8.10 Making Detailed Set-
tings for Input Modules" (
39
Auto balance failed.
Check whether a sensor is in an uncharged
state, and that it is connected correctly.
"8.10.4 Settings for the
8969 Strain Unit" (
40
Voltage Sag triggering is disabled. (Val-
id time base range: 20
μ
s/div to 50 ms/
div)
Voltage Sag triggering can be used only when
the time base is between
20
μ
s/div to 50 ms/div
"9.3.1 Analog Trigger Set-
tings and Types" (
41
Channels that are not measurable will
be assigned to X-Y.
Channel with specification exceeding [Used
Ch] is selected. Change channel selection.
Use (Extending the
Recording Length)
42
There is not enough data in the memo-
ry.
Measure the data portion required for calcula-
tion.
43
Stopped.
-
-
44
Current clamp/sensor was recognized.
-
-
45
Current clamp/sensor was disconnect-
ed.
Check the connection of the current clamp/
sensor.
-
46
Channels may no longer be able to be
used by frequency units due to logic
use.
If logic channels A, B, C, and D of the machine
are used, the respective frequency units at
channels 1, 2, 3, and 4 cannot be used.
47
:This unit cannot be used because it is
not AAF aligned.
Press the SYSTEM key to display the
[Init]
sheet.
Execute the
[System Information]
and check
the software items.
If
[Not AAF aligned]
is displayed, send the
unit for repairs.
-
48
There are units on which AAF cannot be
set to On because of lack of adjustment.
50
Roll Mode is not available.
The Roll Mode cannot be used when Overlay
is enabled.
"8.2 Displaying Waveforms
During Recording (Roll
Mode)" (
51
Cannot use. (Pre Trigger)
The pre-trigger function is not available when
using external sampling.
"Chapter 16 External Con-
trol" (
52
Cannot use .(Roll Mode, Memory Divi-
sion)
If the waveform calculation function is used,
these functions cannot be used.
53
Cannot use. (Roll Mode, Memory Divi-
sion, Wave Calculation)
If one function is used, the other functions
cannot be used.
"8.2 Displaying Waveforms
During Recording (Roll
Mode)" (
Displayed Warnings
Msg
No.
Message
Remedial Action
Reference
Summary of Contents for 8847
Page 1: ...MEMORY HiCORDER Instruction Manual 8847 December 2010 Revised edition 4 8847A981 04 10 12H ...
Page 2: ......
Page 18: ...Operating Precautions 8 ...
Page 26: ...1 4 Basic Key Operations 16 ...
Page 74: ...4 4 Waveform Observation 64 ...
Page 98: ...5 5 Managing Files 88 ...
Page 110: ...6 6 Miscellaneous Printing Functions 100 ...
Page 126: ...7 8 Seeing Block Waveforms 116 ...
Page 160: ...8 10 Making Detailed Settings for Input Modules 150 ...
Page 198: ...10 6 Numerical Calculation Type and Description 188 ...
Page 296: ...15 7 Controlling the Instrument with Command Communications LAN USB 286 ...
Page 304: ...16 2 External I O 294 ...
Page 362: ...Appendix 5 FFT Definitions A28 ...
Page 368: ...Index Index 6 ...
Page 369: ......
Page 370: ......
Page 371: ......