2.2 Connecting Cords
24
Applicable Input Modules
• 8969 Strain Unit
Use to connect: Sensor
• Strain Gauge Transducer (Not avail-
able from Hioki)
• 9769 Conversion Cable
Connecting using a 9769
Conversion Cable
Using a Strain Gauge to Measure Vibration or Displacement (Strain)
Connect Model 9769 Conversion Cable to
the input module jack.
Attach to the measurement object
Connecting the 9769
1
Connect the 9769 to a terminal on the
input module.
The orange section of the 9769 must
face up.
2
Connect the strain gauge transducer to
the conversion cable.
3
Attach to the measurement object.
Example: Connecting the 9769 Conversion Cable with the supplied conversion cable
Required item:
9769 Conversion Cable, Strain Gauge Converter
Connect to input module's terminal
Connect to the Strain Gauge Converter
3
2
1
Bridge voltage:
2 V is supplied.
Connector pinout of 9769 on sensor side
Metal shell is connected to GND of the 8969.
Connector Pinout of the 8969
(1 is on left when unit top side is facing up)
Pin No.
Description
1
2
SENSE+
3
INPUT+
4
INPUT-
5
BRIDGE-
6
SENSE-
7
FLOATING COMMON
Pin Mark
Description
A
B
INPUT-
C
BRIDGE-
D
INPUT+
E
FLOATING COMMON
F, G
N.C.
1 2 3 4 5 6 7
Summary of Contents for 8847
Page 1: ...MEMORY HiCORDER Instruction Manual 8847 December 2010 Revised edition 4 8847A981 04 10 12H ...
Page 2: ......
Page 18: ...Operating Precautions 8 ...
Page 26: ...1 4 Basic Key Operations 16 ...
Page 74: ...4 4 Waveform Observation 64 ...
Page 98: ...5 5 Managing Files 88 ...
Page 110: ...6 6 Miscellaneous Printing Functions 100 ...
Page 126: ...7 8 Seeing Block Waveforms 116 ...
Page 160: ...8 10 Making Detailed Settings for Input Modules 150 ...
Page 198: ...10 6 Numerical Calculation Type and Description 188 ...
Page 296: ...15 7 Controlling the Instrument with Command Communications LAN USB 286 ...
Page 304: ...16 2 External I O 294 ...
Page 362: ...Appendix 5 FFT Definitions A28 ...
Page 368: ...Index Index 6 ...
Page 369: ......
Page 370: ......
Page 371: ......