background image

Introduction

Thank you for purchasing the HIOKI "9264-01/02 WIRING

ADAPTER." To obtain maximum performance from the prod-

uct, please read this manual first, and keep it handy for future

reference

The Wiring Adapter are specially designed to facilitate con-

nections to three-phase 3- and 4-wire systems using the 3196

POWER QUALITY ANALYZER.

Model 9264-01 WIRING ADAPTER

:

For three-phase 3-wire (3P3W3M)

Only three voltage cords need to be connected to the circuit,

instead of six.

Model 9264-02 WIRING ADAPTER

:

For three-phase 4-wire (3P4W)

Only four voltage cords need to be connected to the circuit,

instead of six.

For safety and maintenance, read the Quick Start Manual of

the 3196 POWER QUALITY ANALYZER.

Symbols

Overview

Specifications

Compatible Model

3196

Operating environment Indoors, altitude up to 2000 m (6562-ft.)
Storage temperature 

and humidity

-20 to 50

°

C

 (-4 to 122

°

F), 

80% RH or less (non-condensating)

Operating tempera-

ture and humidity

0 to 40

°

C (32 to 104

°

F), 

80% RH or less (non-condensating)

Maximum input voltage 780 Vrms, 1103 V peak (between A and 

B, B and C, C and A)

Maximum rated volt-

age to earth

600 Vrms (9264-02: between A and N, B 

and N, C and N)

Dielectric strength 

(50/60 Hz, for one 

minute)

5.55 kVrms for one minute (current sensi-

tivity 1 mA) (Between voltage input termi-

nals and case)

Applicable Standards 

(Safety)

EN61010-1:2001

Voltage Input: Pollution Degree 2, 

Measurement Category III (anticipated 

transient overvoltage 6000V)

Dimensions

Approx. 125W x 80H x 32D mm 

(4.92"W x 3.15"H x 1.26"D)

(Cable length: 152 mm, 5.98")

Mass

Approx. 180 g (6.3 oz)

9264-01, 9264-02

WIRING ADAPTER

INSTRUCTION MANUAL

November 2003  Revised edition 1  Printed in Japan

9264A980-01  03-11H

HEAD OFFICE

81 Koizumi, Ueda, Nagano 386-1192, Japan

TEL +81-268-28-0562 / FAX +81-268-28-0568

E-mail: os-com@hioki.co.jp  URL http://www.hioki.co.jp/

HIOKI USA CORPORATION

6 Corporate Drive, Cranbury, NJ 08512, USA

TEL +1-609-409-9109 / FAX +1-609-409-9108

Before Use

In the manual, the 

 symbol indicates particularly im-

portant information that the user should read before using 

the product.

The 

 symbol printed on the product indicates that the 

user should refer to a corresponding topic in the manual 

(marked with the 

 symbol) before using the relevant 

function.
Indicates that incorrect operation presents an extreme 

hazard that could result in serious injury or death to the 

user.

Connecting a Wiring Adapter

• The maximum input voltage is 780 Vrms, 1103 V peak.

Do not measure the voltage exceeding that. It causes

the personal injury as well as damages the product.

• The maximum rated voltage to earth is 600 Vrms. 

Do not measure the voltage to earth exceeding that. It

causes the personal injury as well as damages the

product.

• To avoid electric shock, turn the 3196 off when con-

necting a wiring adapter. 

3196

9264-01

Connect all six output

leads from the Wiring

Adapter to the voltage

input terminals with the

corresponding colors on

the 3196.

Insert plugs all

the way in.

A
B
C

U-  U+   U-  U+   U-  U+

U INPUT

CH1

CH2

CH3

CH

LO

A

D

Bl

ue

A,B,C:Line

Ye

llo

w

Red

S

O

U

R

C

E

A

B

C

Red

Ye

llo

w

Blue

Bl

ack

Bl

ack

Bl

ack

Three-phase 3-wire 

(3P3W3M)

A
B
C

U INPUT

CH1

CH2

CH3

CH

Red

N

A,B,C:Line, N:Neutral

Blue

Black

Ye

llo

w

U-  U+   U-  U+   U-  U+

S

O

U

R

C

E

LO

A

D

A

B

C

N

Red

Ye

llow

Bl

ue

Bla

ck

Bla

ck

Bla

ck

Three-phase 4-wire 

(3P4W)

9264-02 WIRING ADAPTER

9264-01 WIRING ADAPTER

U-  U+ U-  U+ U-  U+

CH1

CH2

CH3

A

B

C

N

U-  U+ U-  U+ U-  U+

CH1

CH2

CH3

A

B

C