Interfaces
14
Network Addresses : 192.168.1.0/24 (Private IP addresses)
Subnet Mask
: 255.255.255.0
PC (The settings are made manually)
IP Address
: 192.168.1.1
For a network as shown above, set the PC and the instrument's Comm screen as shown
next.
When connecting one PC and one instrument by the conversion connector of the 9642 LAN
CABLE, the IP address can be specified arbitrarily, but there is no problem with using a pri-
vate IP address.
Set the instrument as shown below, so that every instrument has a unique host name and
IP address.
LAN Connection Examples
Connecting one PC and one instrument with a 9642 LAN CABLE.
PC Settings
IP Address
: 192.168.1.1
Subnet Mask: 255.255.255.0
The network settings on the PC are made in
the
[Network]
dialog box. To get to this dialog
box, double-click the
[Network]
icon in
[Control
Panels]
.
Instrument Settings
Host Name
: LOGGER
DHCP
: OFF
IP Address
: 192.168.1.2
Subnet Mask
: 255.255.255.0
Port Number
: 880X
Gateway
: OFF
DNS
: OFF