9.3 Troubleshooting
93
3
9
Chapter 9 M
aintenanc
e and Service
Logger settings cannot be
changed.
Dead battery.
When the
battery indicator ap-
pears, settings cannot be changed
(but only displayed). Replace the bat-
tery.
See:
"2.1 Installing (or Replacing)
the Battery" ( p.17)
How can the logger's mem-
ory be erased?
−
Logger memory can be erased using
the LR5000 Utility Program.
See:
"Other Settings on the Logger
Settings Screen" ( p.38)
Note that data recorded prior to the
last recording is automatically erased
whenever recording starts.
(The logger retains the data from both
current and most recent prior record-
ing operation.)
See:
"4.3 Starting and Stopping
Recording" ( p.42)
How can recorded values
be reorganized?
−
Enable scaling.
See:
Scaling settings can be made before
recording.
See:
"Scaling (set as needed)" ( p.36)
Recorded data has disap-
peared.
Recording was restarted after
stopping.
Note that if recording is accidentally
restarted after stopping, data record-
ed prior to the last recording is auto-
matically erased. (The logger retains
the data from both current and most
recent prior recording operations.)
Before requesting repairs
Problem Symptom
Probable Causes
Remedies and References
HIOKI LR5011B980-04
Summary of Contents for LR5011
Page 2: ...HIOKI LR5011B980 04...
Page 6: ...Contents iv HIOKI LR5011B980 04...
Page 66: ...4 8 Printing Recorded Data 60 HIOKI LR5011B980 04...
Page 82: ...6 4 Extracting Data 76 HIOKI LR5011B980 04...
Page 102: ...9 4 Error Displays 96 HIOKI LR5011B980 04...
Page 108: ...Index Index 4 HIOKI LR5011B980 04...
Page 109: ...HIOKI LR5011B980 04...
Page 110: ...HIOKI LR5011B980 04...
Page 111: ...HIOKI LR5011B980 04...
Page 112: ...HIOKI LR5011B980 04...