8.1 Measurement Specifications
81
3
8
Cha
p
te
r 8
Specif
icat
ions
Specifications
Chapter 8
8.1 Measurement Specifications
Sensor
External temperature sensor 1channel (Thermistor)
Measurement ranges
• Temperature: -40.0
°
C to 180.0
°
C (-40.0
°
F to 356.0
°
F)
Note 1: Measurement range is limited according to sensor type.
Note 2: "UF" or "OF" indicates out-of-range measurement.
Measurement accuracy
( sensor)
• Temperature:
between -40.0
°
C (-40.0
°
F) and 0.0
°
C (32.0
°
F)
: ±1.0
°
C (±1.8
°
F)
between 0.0
°
C (32.0
°
F) and 35.0
°
C (95.0
°
F)
: ±0.5
°
C (±0.9
°
F)
between 35.0
°
C (95.0
°
F) and 70.0
°
C (158.0
°
F)
: ±1.0
°
C (±1.8
°
F)
between 70.0
°
C (158.0
°
F) and 120.0
°
C (248.0
°
F) : ±2.0
°
C (±3.6
°
F)
120.0 to 180.0
°
C (248.0 to 356.0
°
F)
: ±5.0
°
C (±9.0
°
F)
Accuracy guarantee for
temperature and humidity
• Temperature: -20.0
°
C to 70.0
°
C (-4.0
°
F to 158.0
°
F) (logger)
• Humidity: 80%RH or less (logger) non-condensating
Product warranty period 3 years
Guaranteed
accuracy period
1 year
±5
±4
±3
±2
±1
0
0
-40
35
70
120
180
Measurement temperature [
°
C]
Measurement
error [
°
C]
HIOKI LR5011B980-04
Summary of Contents for LR5011
Page 2: ...HIOKI LR5011B980 04...
Page 6: ...Contents iv HIOKI LR5011B980 04...
Page 66: ...4 8 Printing Recorded Data 60 HIOKI LR5011B980 04...
Page 82: ...6 4 Extracting Data 76 HIOKI LR5011B980 04...
Page 102: ...9 4 Error Displays 96 HIOKI LR5011B980 04...
Page 108: ...Index Index 4 HIOKI LR5011B980 04...
Page 109: ...HIOKI LR5011B980 04...
Page 110: ...HIOKI LR5011B980 04...
Page 111: ...HIOKI LR5011B980 04...
Page 112: ...HIOKI LR5011B980 04...