2.2 Connecting the Connection Cable
22
Connect a connection cable to the logger's sensor jacks.
Required Items: Hioki LR9801 Connection Cable
LR9801 Connection Cable
2.2
Connecting the Connection Cable
• To avoid breaking the sensor, do not bend or pull it.
• Avoid stepping on or pinching cables, which could damage the cable
insulation.
• A fuse is built into the logger to protect the internal circuit from excess
power input.
Do not connect it to a location with more than 2.5 V as
the fuse will melt and measurements cannot be taken.
Connection Method
Align the triangles on the plug body and in front of
the connection cable jacks, and insert the plug
securely.
Correct values are not displayed unless the plug is
inserted all the way in.
If values are not displayed correctly even when the
plug is inserted properly, the logger or cable may
be damaged. Repair may be necessary.
See:
Connection Cable
Cable length: Approx. 1000 mm
White Wire: Current Input (+)
Black Wire: Current Input (-)
Current Input
(+)