3.2 Making Settings on the Logger
33
3
3
Chapter 3 Set
tings
Stop Method Setting (for when memory becomes full)
1
Press the
SET
button to display the stop
method setting.
(The
[ENDLESS]
indicator
appears, and the setting blinks.)
2
Press the
(+)
and
(-)
buttons to select
[ON]
or
[OFF]
.
3
Press the
SET
button to accept the set-
ting.
(The recording mode setting is displayed.)
Setting Options
Description
OFF
Recording stops when memory becomes full (One-Time Recording).
ON (Default)
The oldest data is overwritten when memory is full (Endless Recording).
When memory becomes full during one-time recording, the recorded
data count appears as follows.
When memory becomes full during endless recording, the recorded
data count (equal to the memory capacity) remains constant.
(the Measurement display shows channel mea-
surement value and recorded data count)
(instantaneous value recording display)
(statistical value recording display)