9.4 Error Display
94
The display appears as follows when an error occurs on the logger.
9.4
Error Display
Logger Error Displays
Error Displays
Meaning
Remedies and References
Calibration data error:
A fault occurred with the internal
calibration data.
Inspection and repair is required.
Please contact your dealer or Hioki
representative.
See:
Microcomputer error:
A fault occurred in microcomputer
ROM/RAM.
Data recording error:
A fault occurred in recording data
or accessing settings.
Battery voltage is too low for nor-
mal logger operation.
Replace the battery.
See:
"2.1 Installing (or Replacing)
the Battery" (p.19)
or
A measured value is out of range.
Out-of-range values cannot be dis-
played.
[OF]
or
[UF]
is displayed when this
data is imported by the LR5000 Utility
Program.
LR5000 Utility Program Error Displays
Error Displays
Meaning
Remedies and References
OF
A measured value is out of range.
Out-of-range values cannot be dis-
played.
UF