4.1 Collecting Recorded Data of Logger
37
3
4
Chapter 4 Collec
tin
g an
d Br
o
w
sing
Dat
a
• In the case of a logger for which data was previously collected (logger
with same serial number), the data is saved to the location where the
previous data was saved. If data exists in both the collector memory
and SD memory card, the save destination becomes the SD memory
card.
• Data can be collected without stopping measurement on the logger.
Data is collected up until the point in time when data collection was
performed. However, the time required to perform data collection is
longer than when measurement is stopped.
What should I do if an error message appears?
See:
"When attempting to collect recorded data:" (p.130)
What happens if collection is performed once during recording and
then again after recording is finished?
Select whether to collect only uncollected data or all data for the collection
method in step "5" (p.35).
What should be done if communication is interrupted during col-
lection?
Check whether or not the IR port is scratched or dirty. Check the connection to
the logger, and then perform collection again. The uncollected data will be col-
lected.