7.2 Setting the Collector from the LR5000 Utility Program
80
Alarm Thresholds (set as needed)
Set the upper and lower alarm threshold values.
When a measurement is outside of the specified area, the
[AL]
(alarm) indicator is
displayed on the logger.
Click the
[Save]
button to save your settings.
(The display returns to the Logger Settings screen.)
Note: If you click the
[Cancel]
button without saving the settings, the display still returns to the
Logger Settings screen.
Note: Alarm judgment is performed at every recording interval during instantaneous recording,
and once per second during statistical recording.
Note: Alarm judgment is performed using measurement values with a larger number of digits
than the values (4 digits) indicated in the Data Logger's display.
Note: The
[AL]
indicator appears when the measured value is out of range (OF/UF dis-
played), and when a sensor anomaly occurs (- - - - displayed).
Upper and lower thresholds
Enter numerical values between -9999 and 9999 (up to six digits).
When scaling is enabled, enter these values as scaled results.
Enable alarm judgment function
Select this check box to enable the alarm.