84
Setting Measurement Conditions
Scaling setting example
When integrating the number of pulses with a 50,000 pulse/kWh power meter
connected:
Scaling decimal
Unit kWh
1 kWh = 50,000 pulses
When integrating the number of pulses with a 10 L/pulse flow meter connected:
Scaling decimal
Unit L
1 pulse = 10 L
Setting the displayed digit under decimal point
The measurement value is displayed with the decimal point fixed to the specified
digit.
Settings:
AUTO
Displays a 4-digit value (0.000 to ±9,999). The decimal point position is
changed as needed.
0 Digit/1 Digit
/2 Digit/3 Digit
The decimal point is fixed to the specified digit.
Setting the auxiliary unit
The value obtained when the measurement value is multiplied by the constant shown
in the following table is displayed in the wireless logger screen. (The measurement
value is not affected.)
Settings:
μ (micro)
×10
6
m (milli)
×10
3
-
×1
k (kilo)
×10
-3
M (mega)
×10
-6
Reset
The conversion parameter (conversion ratio, offset, input, output) is restored to the
default setting.
www.
GlobalTestSupply
.com
Find Quality Products Online at:
sales@GlobalTestSupply.com