11.2 Configuring Pulse Settings
153
Cha
p
te
r 1
1
Using Pulse
Input
and Output
11
When using the pulse I/O terminals, you must configure the instrument’s pulse set-
tings.
11.2 Configuring Pulse Settings
1
Press the
key to display the
[SET 8/8, PULSE]
screen.
2
Configure the settings as desired.
Pulese input
Filter ON (mechanical contact use):
Frequencies of 25 Hz or less (with high and low intervals at least 20 ms in du-
ration)
Filter OFF (electronic contact use):
Frequencies of 5 kHz or less (with high and low intervals at least 100
μ
s in
duration)
Pulse output
FILTER
ON/OFF
SCALING
0.001 to 1.000 to
100.000
AUX UNIT
p/n/
(u)/m/none/k
/M/G/T
UNIT
Up to 5 byte characters
OUTPUT RATE
OFF/1 Wh/10 Wh/100 Wh/1 kWh/10 kWh/100 kWh/1000 kWh
PULSE WIDTH
Fixed to 100 ms (cannot be changed)
Because the active energy value is updated every second, the
instrument can generate pulse output at a maximum rate of 1
pulse per second. Set the pulse output rate so that it is greater
than the active energy consumed in 1 second.
Example:
If the maximum active energy consumption in 1 second is 150 Wh,
the pulse output rate should be set to a value that is greater than
or equal to 1 kWh.
Summary of Contents for PW3360-20
Page 2: ......
Page 66: ...3 8 Verifying Correct Wiring Wiring Check 58 ...
Page 104: ...7 2 Settings That Can Be Added to Quick Set Settings 96 ...
Page 120: ...8 8 Formatting the SD Memory Card or Internal Memory 112 ...
Page 166: ...11 4 Outputting a Pulse Signal 158 ...
Page 210: ...13 4 Disposing of the Instrument 202 ...
Page 218: ...A8 ...
Page 222: ...Index Index 4 ...
Page 223: ......
Page 224: ......