35
Query measurement value in memory (range specification)
Syntax
Query
:MEMory:RANGe?
<1
~
255>,
<Start memory number>,<End memory number>
Response
< Return measurement value and judgment >
Description
Return measurement value and judgment
Set the data you wish to query as a value from 1 to 255. For example, to query measured
values only you would set the value 2. To query measured values along with judgment
results, you would set 2 and 4 for a total of 6.
<Start memory number> = 1 to 999
<End memory number> = 1 to 999
If there is no data stored in its memory, the instrument will return “0.” Specifying a memory
number in which no data is stored will result in an execution error.
128
64
32
16
8
4
2
1
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
Voltage
check
result
Contact
check
result
Humidity Tempereture
Voltage
Monitor
Judgment
value
Measurement
value
Unused
Example
:MEMory:RANGe? 6,10,20
6.33802E-12,HI, 6.33533E-12,HI, 6.33833E-12,HI,……
Set and Query memory state
Syntax
command
:MEMory:STATe
<ON/OFF>
Query
:MEMory:STATe?
Response
<ON/OFF>
Description
ON … save measurement value
OFF … no save
Example
:MEMory:STATe
ON
Set measurement value to save
:MEMory:STATe?
ON
Measurement value has been save
Query number of measurement value
Syntax
command
:MEMory:COUNt?
Response
<0
~
50(NR1)>
Description
Return number of measuremet value
Example
:MEMory:COUNt?
23
Mumber of measuremet value has been 23
Set Memory clear
Syntax
command
:MEMory:CLEar
Description
Clear all measurement value in memory
Example
:MEMory:CLEar
Clear all measurement value in memory
Note
Can't return the panel data once delete