20
(3)
Voltage Monitor
Query Voltage Monitor value
Syntax
Query
:MEASure:MONItor?
Response
<Voltage monitor (NR2)>
Description
Return the Voltage monitor value
Example
:MEASure:MONItor?
500.2
Voltage monitor value has been 500.2V
(4)
Tempureture and Humidity value
Query Tempureture value
Syntax
Query
:MEASure:TEMPerature?
Response
<Tempureture(
℃
)(NR2)>
Description
Return tempureture value
When Z2011 has not set, return tempreture value has been 99.99
Example
:MEASure:TEMPerature?
23.45
Tempureture has been 23.45
℃
Query Humidity value
Syntax
Query
:MEASure:HUMidity?
Response
<Humidity(%rh)(NR2)>
Description
Return humidity value
When Z2011 has not set, return humidity value has been 99.99
Example
:MEASure:HUMidity?
50.1
Humidity has been 50.1 %rh
(5)
Measured value display mode
Set and Query Measured value display mode
Syntax
command
:MEASure:MODE
<R/A/RS/RV/RL>
Query
:MEASure:MODE?
Response
<R/A/RS/RV/RL>
Description
R … Resistance
A … Current
RS … Surface resistivity
RV … Volume resistivity
RL … Liquid volume resistivity
Example
:MEASure:MODE R
Set Resistance mode
:MEASure:MODE?
R
Measured value mode has been Resistance