Operational Instructions and Function Adjustment
Mode
Select the display mode to be adjusted. The following table lists the initial values of all modes.
Mode
Brightness (cd/m
2
)
Gamma Value
DICOM300
300
DICOM
DICOM350
350
DICOM
DICOM400
400
DICOM
DICOM450
450
DICOM
DICOM500
500
DICOM
DIC0M550
550
DICOM
DICOM600
600
DICOM
DIC0M650
650
DICOM
DICOM700
700
DICOM
DICOM User
500
DICOM
Custom
300
Gamma 2.2
ALC
500
DICOM
Note
After you select a mode and adjust its value, the LCD monitor displays content based on the
adjusted value.
Brightness
You can use this button to set the current monitor brightness. If the current mode is DICOM300,
DICOM350, DICOM400, DICOM450, DICOM500, DICOM550, DICOM600, DICOM650, or
DICOM700, the system automatically enters the DICOM mode after the brightness is adjusted.
Contrast
Set
Contrast
to adjust the contrast ratio in the current display mode. The value range is 0 to 100.
The default value is 50.
Note
This item can be adjusted only in Custom mode.
15