background image

 

 

Overview 

 

Thank you for purchasing Hisense LCD monitor, which complies with the DICOM Part 14 standard 
and is applicable to DSA, MRI, DR, CR, CT, and PET medical imaging. It is intended for 
professionally trained medical staff and provides the image viewing and medical diagnostic functions. 

 

Features 

 

Environmental friendliness: 

The LCD monitor is equipped with an energy-efficient LED backlight LCD panel with a long 
service life. The backlight is driven by low voltage, which eliminates the hazardous radiation to 
the body and environment. 

 

Front-facing sensor calibration: 

The LCD monitor supports gray scale inspection and automatic calibration by using the 
front-facing sensor and quality control software. 

 

Digital ports: 

It has multiple input interfaces (two DVI-Ds, one HDMI, and two DPs). It supports the following 
functions: DP daisy chain, simultaneous display of two signal sources, built-in Picture by Picture 
(PBP), Picture in Picture (PIP), multi-channel independent GAMMA correction. It also supports 
adjusting the image mode of each window independently to match the characteristics of their 
respective signal sources. 

 

Human sensor: 

It automatically identifies whether the operator is in front of the monitor and automatically makes 
the monitor enter and awake from the sleep status to save electricity and prolong the display 
service life. 

 

HD display and processing technology: 

With a resolution of 3280x2048, the monitor can accurately display images in each mode, without 
any blur or distortion. 

 

 

 

2

 

Summary of Contents for HMD6C30S

Page 1: ...HMD6C30S User Guide Read the User Guide thoroughly before using Hisense LCD monitor and keep it properly for future reference RSAG2 025 7172SS B Version Hisense website www hisense com...

Page 2: ...the User Guide Hisense national customer service hotline 4009940707 Hisense service quality supervision tel 0532 83091111 Only the repair personnel are allowed to repair the LCD monitor and replace it...

Page 3: ...and maintenance when the LCD monitor is energized Disconnect the LCD monitor power cable whenever possible Description and Maintenance of the LCD monitor The screen is a key component of the LCD moni...

Page 4: ...LCD monitor is abnormal contact Hisense or an authorized distributor Ensure that no sundries exist in the surrounding area of mains disconnection devices such as the switch power plug and appliance co...

Page 5: ...or and Input Ports 7 Hanging Arm Installation 8 PC Connection 9 Operational Instructions and Function Adjustment 10 Menu Adjustment 10 Main Menu Display 10 Setting and Adjustment 11 User Menu Display...

Page 6: ...ing the front facing sensor and quality control software Digital ports It has multiple input interfaces two DVI Ds one HDMI and two DPs It supports the following functions DP daisy chain simultaneous...

Page 7: ...W input signal connector DVI D connector DisplayPort connector HDMI connector Output signal connector DisplayPort connector Dimensions W x H x D System with the base 700 x 516 626 x 274 mm System wit...

Page 8: ...software Graphics card 1 Image clip 1 Quick installation manual 1 Note The accessory list is for reference only If any component is lost or damaged promptly contact Hisense or the specified distribut...

Page 9: ...highest position CAUTION When adjusting the height of the monitor hold the left and right sides of the monitor with both hands Do not put your hands directly under the monitor to avoid pinching your h...

Page 10: ...cklight brightness increases decreases Downward Back It is used to return to the previous menu When no menu is available it is used to call out the Image Enhancing function Display mode The display mo...

Page 11: ...monitor or remove the power plug to save the energy Rear View of the LCD Monitor and Input Ports Port Function Power port Connected to the power adapter DEBUG 1 2 3 and 4 It is used for product debug...

Page 12: ...T 819 1 2000 M4x14 When using the hanging arm or base install it based on the following inclination of the monitor 15 upwards and 5 downwards Connect cables after the hanging arm or base is installed...

Page 13: ...ix the connection The following takes the DVI and DP signal cables as an example 2 Insert the power adapter to the power socket and the power connector on the monitor 3 After the power is turned on th...

Page 14: ...ngle PBP PIP Signal Input Single DVI1 DVI2 DP1 DP2 and HDMI PBP PIP Image Settings Windows Window 1 Window 2 Image mode DICOM CT MRI DR DSA Video Text Hybrid GAMMA User Zoom Full Screen Scale Screen D...

Page 15: ...Extra Function Multi Window Settings Three Four Back to original mode Logo Set On Off Factory Reset Confirm and Cancel Setting and Adjustment Press on the monitor to call out the User menu Press or t...

Page 16: ...to the Signal Input menu 5 Press to return to the User menu User Menu Image Settings You can set different signal sources for each window in different display modes Procedure 1 In the User menu press...

Page 17: ...00 550 and 600 unit nit gears User Menu Calibration Settings In Single mode this function is used to calibrate and validate the DICOM curve Procedure 1 In the User menu press or to select calibration...

Page 18: ...sor to On or Off 6 In the Human Sensor menu you can set the human sensor to On or Off 7 In the Auto Awake menu you can set whether to awake the monitor when a DP1 communication signal is coming in the...

Page 19: ...y the LCD monitor Disconnect and then reconnect the signal cable or restart the PC The image is blurred the screen displays interference stripes or characters show prominent edge colors Check whether...

Page 20: ...es emitted by power supply equipment may affect the use of the LCD monitor and damage it Install the LCD monitor in an environment free of electromagnetic wave interference If the LCD monitor generate...

Page 21: ...ing harmful substances Environmental friendly service life The environmental friendly service life identifier printed on the LCD monitor s nameplate and storage box indicates that the LCD monitor has...

Page 22: ...idents incorrect installation natural disasters or improper maintenance by a third party other than Hisense and authorized distributors The product ages because it has been used with brightness above...

Page 23: ...itor s brightness and its input voltage is non linear A smaller gamma value generates an image with lower contrast whereas a greater gamma value generates an image with higher contrast In Plane Switch...

Page 24: ...Co Ltd Address North 3F Outsourcing Center of Software Park No 169 Songling Road Laoshan District Qingdao P R China Hisense national customer service hotline 4009940707 Hisense Group service supervis...

Reviews: