-20-
MN1-5008 Rev. 7
4-4. Connecting to the ultrasound diagnostic instrument
The lock lever of the connector moves over the range shown in the figure
at right.
Align the ○ mark with the LOCK or RELEASE position and lock or
release the electronic probe connecting socket of the diagnostic instrument
( probe connector ).
Connect the scanner to the probe connector by following the procedure
below.
• Connection procedure
The scanner is connected when in one of the following states.
• The power switch is set to OFF.
• The image displayed on the ultrasound diagnostic instrument is
frozen.
Before inserting the scanner into the probe connector, check that the
connector pins are not bent.
1. Turn the connector lock lever to align the ○ mark on the lever with the
RELEASE position.
2. Firmly insert the connector into the probe connector on the ultrasound
diagnostic instrument.
3. Turn the lock lever clockwise by 1/4 turn until the ○ mark is aligned
with the LOCK position.
4. Check that the connector is firmly inserted into the probe connector on
the instrument.
5. Connect the cable for driving to the upper part of the connector box.
This completes connection of the scanner.
Caution
If there is resistance when trying to turn the lock lever when connecting the connector, do not
forcibly try to connect it. Instead, correctly perform the steps for connecting the connector and
firmly insert it into the probe connector.
Forcibly turning the lever may damage the connector and the probe connector.
Lock lever
Movable
range
○ mark
LOCK
RELEASE
RELEASE
LOCK
Small connector
Summary of Contents for ASU-1003
Page 2: ...MN1 5008 Rev 7 ii ...
Page 6: ...MN1 5008 Rev 7 vi ...
Page 14: ... 8 MN1 5008 Rev 7 Label A 2 Storage case Label B ...
Page 22: ... 16 MN1 5008 Rev 7 ...
Page 34: ... 28 MN1 5008 Rev 7 ...
Page 36: ... 30 MN1 5008 Rev 7 ...
Page 38: ... 32 MN1 5008 Rev 7 ...
Page 42: ... 36 MN1 5008 Rev 7 ...