-23-
MN1-5008 Rev. 7
Caution
Do not place the insertion portion and handle in any liquids
beyond the range shown in the figure right.
The connector which liquid has intruded can cause the
malfunction of the probe and the ultrasound diagnostic
instrument. In this case immediately stop use and contact one
of our offices and/or distributor's offices listed on the back
cover.
5. Washing, Disinfection and Sterilization
Warning
Wear protective gloves and other protective gear during washing, disinfection and sterilization.
Handling of the scanner with bare hands before disinfection or sterilization can result in an
infection.
After soaking in cleaning agents, thoroughly wash the scanner with running water.
Residual cleaning agents can cause an adverse reaction on the bodies of the operator or patient.
After chemical disinfection and sterilization,
thoroughly wash the scanner with sterilized water.
Residual chemicals can cause an adverse reaction on the bodies of the operator or patient.
Perform aeration completely after gas disinfection.
Residual gas can cause an adverse reaction on the bodies of the operator or patient.
Do not wash, disinfect or sterilize using procedures other than those specified in this manual.
Infection could result due to incomplete washing disinfection or sterilization. It can also result in
damage to the scanner or reduced performance. Using the damaged scanner has occurred can result
in an electric shock and injury to the operator or patient.
The probe cannot withstand autoclave sterilization or boiling and other types of sterilization at
temperatures exceeding 60°C (140°F).
For details on the usage conditions of chemicals and sterilization procedures, refer to the
documentation supplied with the respective chemical or sterilization equipment.
Infection could result due to incomplete disinfection or sterilization.
This could also cause deterioration of the scanner.
Water or chemical solution
150mm or less
Summary of Contents for ASU-1003
Page 2: ...MN1 5008 Rev 7 ii ...
Page 6: ...MN1 5008 Rev 7 vi ...
Page 14: ... 8 MN1 5008 Rev 7 Label A 2 Storage case Label B ...
Page 22: ... 16 MN1 5008 Rev 7 ...
Page 34: ... 28 MN1 5008 Rev 7 ...
Page 36: ... 30 MN1 5008 Rev 7 ...
Page 38: ... 32 MN1 5008 Rev 7 ...
Page 42: ... 36 MN1 5008 Rev 7 ...