-24-
MN1-5008 Rev. 7
5-1. Washing
Wash the insertion portion and handle immediately after use with water or soak in a cleaning agent.
Washing before disinfection and sterilization is very important.
5-1-1. Insertion portion and handle
Applicable cleaning agents
General name
Trade name
Manufacturer
Enzyme cleaning agent
ENZOL™
Practical liquid 0.8V/V%
ADVANCED STERILIZATION PRODUCTS
®
A Johnson & Johnson company
Division of Ethicon, Inc.
Washing procedure
Water rinsing
Rinse the insertion portion and handle with running water.
Wash off blood, mucus and other substances adhering to the insertion
portion and handle using a sponge or gauze.
or
Soaking in
cleaning agent
Minimize the soaking time.
Water rinsing
Sufficiently rinse the insertion portion and handle using running water
to remove all traces of chemicals.
Warning
After soaking in cleaning agents, thoroughly wash the scanner with running water.
Residual cleaning agents can cause an adverse reaction on the bodies of the operator or patient.
5-1-2. Cable and connector
Gently wipe the cable with gauze dipped in ethyl alcohol or water each divided into approximately 20
cm and dry.
Gently clean the connector and other parts of the scanner that must not be soaked in liquid with gauze
dipped in ethyl alcohol and dry.
Note
Wiping the entire length of the cable at once can result in wrinkled surface.
If this occurs, pull the wrinkled part in the opposite direction to undo it.
Summary of Contents for ASU-1003
Page 2: ...MN1 5008 Rev 7 ii ...
Page 6: ...MN1 5008 Rev 7 vi ...
Page 14: ... 8 MN1 5008 Rev 7 Label A 2 Storage case Label B ...
Page 22: ... 16 MN1 5008 Rev 7 ...
Page 34: ... 28 MN1 5008 Rev 7 ...
Page 36: ... 30 MN1 5008 Rev 7 ...
Page 38: ... 32 MN1 5008 Rev 7 ...
Page 42: ... 36 MN1 5008 Rev 7 ...