-25-
Q1E-EP1457
A)
C22P probe
1)
Prepare the disinfectant solution in a tank with cold water (please
follow the instructions of the disinfectant manufacturer regarding
application, concentration, microbiological efficiency, service
life and contact time).
2)
Confirm the concentration of the disinfectant before immersing the
probe. Although Cidex® OPA does not need to be diluted, it is
recommended to use test strips to verify the concentration. The
test strips can indicate whether or not the concentration is above
the Minimum Effective Concentration (MEC). Please also note the
expiration date of the test stripes. Temperature of disinfectant
solution should be minimum 20 °C. The minimum contact time is 5
minutes. If a different disinfectant is used, follow the
manufacturer’s instructions. Please also consider the material
compatibility for the medical device.
3)
Immerse the immersible part of the probe into the disinfectant (see
Fig. 18). Set a clock to insure the recommended contact time which
is 5 minutes.
4)
Rinse the immersible part of the probe with deionized water for
1 minute. (alternatively: immerse the immersible part of the probe
in a tray filled with deionized water (see Fig.18) for 5 min.)
5)
Visually check the outer surface of the probe for leavings of the
disinfectant. If necessary, repeat the rinsing.
B)
Attachment
1)
Prepare the disinfectant solution
as stated in the procedure for the
probe.
2)
Immerse the Attachment into the
disinfectant (see Fig. 18). Set a
clock to insure the recommended
contact time which is 5 minutes.
3)
Rinse the Attachment with deionized
water for 1 minute. (alternatively:
immerse it in a tray filled with
deionized water (see Fig.18) for 5
min.)
4)
Visually check the outer surface of
the Attachment for leavings of the
disinfectant. If necessary, repeat
the rinsing.
Water,
Detergent or
Disinfectant
Fig. 18 Immersion
of the probe and the
Attachment.
Summary of Contents for C22P
Page 48: ... 42 Q1E EP1457 Unit mm Fig 23 Dimensions ...
Page 49: ... 43 Q1E EP1457 Unit mm Fig 24 External view with the attachment ...
Page 50: ... 44 Q1E EP1457 Unit mm Fig 25 Dimensions with the Needle Guide Bracket ...
Page 51: ... 45 E Q1E EP1457 Fig 26 External view with Ultra Pro II Needle Guide cover kit Unit mm ...