Cleaning and Disinfection of EZU-PA7C2
EZU-PA7C2 must be reprocessed after each use. Refer to the
reprocessing instruction in this chapter.
WARNINGS
‐
EZU-PA7C2 is delivered unsterile. Prior to the
first use, reprocess it.
‐
The cavities and the spiral spring of the EZU-PA7C2
require particular attentions during all
processes. The needle guide bracket must be
manually pre-cleaned using ultrasound prior
manual reprocessing as well as automated
reprocessing and must be dismantled into single
parts (shown in Figure 22).
Transportation
before using
EZU-PA7C2 should be packed in a sterile pouch or
container to transport from Central Sterile Supply
Department (CSSD) to an operating room. Be careful
not to damage the sterile pouch or container during
transportation.
Levels of reprocessing requirements: Refer to the chapter 4.