6 - 25
Adjustment > Adjustment Procedure
6-4-3 Burst Level
Preparations:
1) Connect the oscilloscope CH1 to video out.
2) Switch the oscilloscope V-MODE to “CH1”
and TRIGGER SOURCE to “CH1”.
Procedure:
1) Choose BURST LEVEL.
2) Click the EXECUTE button.
3) Click the UP or DOWN button so that burst
l
evel of the waveform is 300mV ± 15 mVp-p.
Click the button at approx. 2-second intervals
while checking any increase or decrease in
burst level.
4) After step 3) is complete, be sure to click the
SAVE button.
Note that clicking the RETURN button will
restore the ADJUST MENU screen to the
status before the adjustment was performed.
5) Click the OK button in dialog to restore the
ADJUST MENU screen.
ADJUST MENU screen
MANUAL ADJUSTMENT PROGRAM for SERVICE STATION
MODEL NAME:
××××
EXECUTE
RETURN
CONNECTION
ADJUST MENU
DATA INITIALIZE
SAMPLING PULSE
AUTO IRIS CONTROL
MATRIX
CHROMA GAIN
AUTO FUCUS
SPOT NOISE
LCD
VIDEO LEVEL
BURST LEVEL
LINEARITY
2
1
BURST LEVEL ADJUSTMENT screen
MANUAL ADJUSTMENT PROGRAM for SERVICE STATION
SAVE
RETURN
BURST LEVEL ADJUSTMENT
UP
DOWN
CONNECTION
4
3
ADJUSTMENT FINISHED dialog
FINISHED
ADJUSTMENT FINISHED.
OK
5
Fig. 6-4-2
M6
M6PP