38 | 108
E-Compact MP-BB3 Series
RF Power Protection:
The reflectometer has accuracy above 10% of the reflected power to react to the equipment reaching or
exceeding 20%. Therefore, accuracy in measures of reflected below 10% relative to direct power are not
guaranteed.
Protection against main voltage variations (Surge):
Each part of the equipment has its independent power supply and all feature the same protection
characteristics.
Surge protection in the grid is carried out by inserting the varistors between the phases and between the
phase and the ground, thereby absorbing the mains voltage peaks and not allowing them to damage the
source.
4.6.6. ALARMS - Front Panel Signaling and Shortcut Keys
FAULT
:
When a fault occurs, the equipment automatically takes protective actions (turn off or reduce power) and
trigger the visual alarm on the front panel (“FAULT” key lit in red). When pressing the FAULT key, the list
of alarms currently occurring appears on the display.
PAST ALARM
:
When there has been a fault and it is not necessarily occurring, “PAST ALARM” will be lit in yellow. When
you press it, it shows the list of alarms that have occurred.
The details of each alarm and where to view it in the display menu: “> Equipment > Alarms”.
ESC
OK
Summary of Contents for EC702MP-BB3
Page 68: ...68 108 E Compact MP BB3 Series VGS Carrier Software Version 1 75 1 70 1 71 1 78 ...
Page 84: ...84 108 E Compact MP BB3 Series B220 BIPHASIC 220VAC 2 PHASE 110VAC GND ...
Page 85: ...85 108 E Compact MP BB3 Series M220 SINGLEPHASE 220VAC 1 PHASE NEUTRAL GND ...
Page 106: ...106 108 E Compact MP BB3 Series Section 8 Attachments ...
Page 108: ...108 108 E Compact MP BB3 Series ...